Proceq 24kHz Transducers for Pundit UPV Instruments
Low frequency travels farther. The 24kHz pair is the choice for very large members and concrete with coarse aggregate, where higher-frequency signals scatter before they reach the receiver.
Specifications
- Frequency: 24kHz
- Wavelength: 154mm
- Maximum aggregate size: about 77mm
- Minimum lateral dimension of test object: about 154mm
For use with the Proceq Pundit Lab+ and Pundit PL-200. For finer resolution on smaller elements, see the 150kHz transducers.