Proceq 150kHz Transducers for Pundit UPV Instruments
- Frequency: 150kHz
- Wavelength: 24.7mm
- Maximum aggregate size: about 12mm
- Minimum lateral dimension: 25mm
Higher frequency means a shorter wavelength and finer resolution. The 150kHz pair suits mortar, fine-aggregate concrete, thin sections and small test objects, where a lower-frequency transducer would not resolve the detail.
Pairs with the Pundit Lab+ UPV instrument. For coarse aggregate and long paths, switch to the 24kHz transducers.