Dry Ice Ultra-Low Temperature Data Logger
The LogTag® Dry Ice “Probe-less” Temperature Logger operates, measures and stores up to 8000 temperature readings in temperature environments ranging from -80°C to +40°C (-112°F to +104°F). Intended for use in transit monitoring of articles stored in packaging incorporating dry ice cooling agents.
The LogTag® Dry Ice Temperature Logger is easily configured for recording conditions including delayed start, sampling interval, number of readings and configuration of conditions to activate the ALERT indicator.
FEATURES:
- Records temperature from +40°C to as low as -80°C
- Up to 8,000 recordings – enough for the longest trip.
- In-transit inspections can be recorded at the push of a button.
- A real time clock provides date/ time stamps for each temperature reading.
- Push-to-start button with optional delay or a specific time & date.
- Comprehensive customization options including alarm settings, sample interval and trip duration.
SPECIFICATIONS:
Product Model
TLSRIL8 – Single use
Sensor Measurement Range
-80°C to +40°C (-112°F to +104°F).
Operating Temperature Range
-80°C to +40°C (-112°F to +104°F).
Storage Temperature Range
-20°C to +40°C (-4°F to +104°F).
Rated Temperature Reading Accuracy
Better than ±1°C (±1.8°F) for -30°C to +20°C (-22°F to +68°F).
Better than ±1.2°C (±2.1°F) for -45°C to -30°C (-49°F to -22°F) and +20°C to +40°C (+68°F to +104°F).
Better than ±1.7°C (±3.1°F) for -80°C to -45°C (-112°F to -49°F).
Actual performance is typically much better than the rated values.
Accuracy figures can be improved by recalibration.
Rated Temperature
Reading Resolution
Less than 0.1°C (0.2°F) for -80°C to 0.0°C (-112°F to +32°F),
Less than 0.2°C (0.4°F) for 0.0°C to +20°C (+32°F to +68°F),
Less than 0.5°C (0.9°F) for +20°C to +40°C (+68°F to +104°F)
LogTag Analyzer® currently displays to one decimal place of °C or °F.
The native resolution is what is stored in the LogTag®.
Sensor Reaction Time
Typically less than 5 minutes (T90) in moving air (1m/s).
Recording Capacity
8031 temperature readings.
53 days @ 10min logging, 80 days @ 15min logging.
Sampling Interval
Configurable from 1 minute to several hours
Logging Start Options
Push button start or specific date & time.
Recording Indication
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
Download Time
Typically less than 5 seconds for full memory (8031 readings),
depending on computer or readout device used.
Environmental
IP65 (roughly equivalent to NEMA 4).
Power Source
3.6V low temperature chemistry Lithium Battery (Fixed).
Battery Life Minimum storage life of 12 months before ‘start’.
Single use version rated for a typical trip of 2 weeks duration at dry ice temperatures.
Real Time Clock
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C
(I.e typically +/- 0.00294 seconds/day/°C).
Size 86mm (3.39″)H x 54.5mm (2.15″)W x 8.6mm (0.34″)T.
Weight 33g (1.17 oz)
Case Material Polycarbonate.