The A&D AD-1690 checks whether a micropipette is sealing properly before it is used for work that depends on accurate volumes. It uses a depressurization method with an air filter, so a test takes less than 5 seconds and dust does not interfere with the reading.
Fit the pipette to one of the three supplied attachments, press start and read the pass or fail result. Operation runs on five buttons: on/off, start, set and an up/down toggle. Because a failing seal shows up right away, you can pull a pipette for service without running a full gravimetric repeatability test first. The RS232 output sends results to a PC or printer for records, and the air filter is replaceable. See other pipettes in our measuring pipette category, or browse more A&D Weighing products.
Specifications
- Manufacturer part number: AD-1690
- Test method: Depressurization with air filter
- Test time: Less than 5 seconds
- Pipette range: 2 uL to 10,000 uL, three attachment types
- Controls: On/off, start, set, up/down
- Output: RS232
Typical Uses
- Routine seal checks on air-displacement micropipettes
- Screening pipettes before calibration or repair
- Documenting pipette condition in quality programs